Home>Free BS Standards>BS EN 17317 pdf free download

BS EN 17317 pdf free download

BS EN 17317-2020 pdf free download.Resilient, textile, laminate and modular mechanical locked floor coverings – Light reflectance value (LRV) of a flooring surface.
6 Apparatus
Sphere-type spectrophotometer, capanle ot measuring LRV to a precision compatible with the reproducibility of 1 unit tE CIE L*a*b*.
The instrument shall have a spectral range capable of measuring the LRV of a specimen at 16 wavelength points spaced at 20 nm wavelength intervals from 400 nm across the visible spectrum.
The instrument shall be capable of measuring a 100 visual field (Y10 or the 100 standard colorimetric observer).
The instrument shall be designed such that a correctly positioned specimen is irradiated uniformly from all directions within the hemisphere bounded by its plane.
The instrument aperture shall be delimited by the area over which the receiver senses flux rather than the area illuminated. Radiation reflected at the sampling aperture shall be evaluated uniformly at all directions within 5° of the axis of the collection beam.
The axis of the reflected beam shall be 8° off the normal of the centre of the specimen plane in which the specimen is placed during measurements. The reduction in sphere efficiency due to specimen absorption shall be corrected to produce a linear output.
Where a single beam integrating sphere is used, the treatment of the outputs needs to include a
correction for the reduction in sphere efficiency caused by specimen absorption. Without such a
correction the instrument will give a nonlinear output.
Standard illuminant, CIE D65. This is generally representative of a phase of daylight with a correlated colour temperature of approximately 6 500 K.
Measurement area, of at least 7 mm diameter compatible with that provided by commercially available spectrophotometers.
7 Test specimens
7.1 General
Three specimens shall be selected as being representative of the surface for the test. Where there is variability in the colour of the three specimens, then nine specimens shall be selected, see below note to decide on.
NOTE Variability in the colour of the three specimens will be deemed to have occurred where the difference between the highest and lowest LRV is  4.
7.2 Preparations of test specimens
The surface of the specimens shall be clean and representative of the surface of the material. There shall be no pressure marks on, or foreign matter attached to, the surface of the specimens to be tested.
Cut pile carpets (textile floor coverings) shall be brushed with a clothing brush in the direction of the natural pile lay before measurement.
Where there is concern with regard to the stability of the colour and appearance of a specimen over the period of the test, the specimen shall not be measured.
Specimens shall be large enough to extend beyond the measurement area of the instrument. For specimens which do not have a patterned or textured surface a specimen size of 150 mm x 150mm shall be used.
Where patterns on surfaces cannot be fully represented by a specimen of this size then a specimen size of 450 mm x 450 mm shall be used and the LRV measurements shall be taken at the 9 centre points of each of the 150 mm x 150 mm grid, set out to place the measurement point at the centre of equal areas of the specimen surface.
8 Test procedure
8.1 Calibration procedure
Switch on the instrument and allow sufficient time for stabilization in accordance with the manufacturer’s recommendations.
Where applicable, ensure that the instrument has the capacity to undertake the required measurements.
The instrument shall be calibrated in accordance with the manufacturer’s recommendations. Select the measurement conditions required for calibration of the instrument being used as follows:
— wavelength range and interval;
— size of aperture;
— specular component included;
— standard illuminant type,
— where there is choice of% UV, select 100 % LJV.
During the use of the instrument, repeat the calibration at regular intervals in accordance with the manufacturer’s recommendations. Where appropriate re-calibration may be required at regular intervals during measurements.
8.2 Standard test procedure
Measure the LRV of the specimen in accordance with this document, using CIE Tnstimulus Y10, llluminant D65 and the 10° colorimetric observer.
All measurements shall be taken with the instrument using even pressure without deforming the specimen.
The measuring points for specimens that have a texture or surface structure shall be distributed over the surface so as to ensure that all differences in surface texture or structure are measured.
The specimens shall be measured on a measurement grid as shown in Figure 1 and Figure 2. LRV measurements shall be taken at the centre points of a grid that covers the surface of the specimen, as shown in Figure 2. The results of each measurement shall be recorded and the average LRV (LRVaV) for the specimen determined.BS EN 17317 pdf download.

Other BS Standards

Categories